SOP for Calibration Of Coating Pan For Rotations Per Minute
Calibration Of Coating Pan For Rotations Per Minute
OBJECTIVE: The objective of this SOP is to provide written guidelines for Calibration Of Coating Pan For Rotations Per Minute |
SCOPE: Scope of this area is limited to Calibration Of Coating Pan For RPM |
RESPONSIBILITY: Operator / Pharmacist |
PROCEDURE:1) Set the rotations per minute of coating pan at 2 rpm,4 rpm,6 rpm,8 rpm & 12 rpm respectively
2) Mark one specific point on pan rim by marker. 3) Measure the time (in seconds) by using calibrated stop watch to complete the set rpm 4) Record such three readings for every rpm setting 5) Calculate the average time (Y) (in seconds) required to complete the set rpm 6) Calculate the rpm of coating pan by using formula, X = set rpm value x 60/ Y 7) Record the observations Frequency: – once in six months Tolerance: – +/- 0.5 rotation per minute
|