SOP for Calibration Of Coating Pan For Rotations Per Minute

0
7584

SOP for Calibration Of Coating Pan For Rotations Per Minute

Calibration Of Coating Pan For Rotations Per Minute

OBJECTIVE: The objective of this SOP is to provide written guidelines for Calibration Of Coating Pan For   Rotations Per Minute
SCOPE: Scope of this area is limited to Calibration Of Coating Pan For RPM
RESPONSIBILITY: Operator / Pharmacist
PROCEDURE:1)      Set the rotations per minute of coating pan at 2 rpm,4 rpm,6 rpm,8 rpm & 12 rpm respectively

2)      Mark one specific point on pan rim by marker.

3)      Measure the time (in seconds) by using calibrated stop watch to complete the set rpm

4)      Record such three readings for every rpm setting

5)      Calculate the average time (Y) (in seconds) required to complete the set rpm

6)      Calculate the rpm of coating pan by using formula,

X = set rpm value x 60/ Y

7) Record the observations

Frequency: – once in six months

Tolerance: – +/- 0.5 rotation per minute